Packaging Materials and Systems for Medical Devices to be Sterilized - Heat and Self-Sealable Pouches and Reels of Paper and Plastic Film Construction (EN 868-5)

This method is used to investigate the strength of the heat seal joints used for heat and self-sealable pouches, and reels for use as packaging of medical devices that are sterilized . The EN 868-5 test method is suitable for use with a range of materials manufactured from paper and plastic film used in packaging.

In principle, the free ends of the test specimen is supported between two clamping jaws and is pulled apart until failure at a constant rate of displacement. This test is to determine the peel characteristics of paper/plastic laminate products, as well as the strength of the heat seal joint for pouches and reel material.

We use screw action grips, pneumatic action grips with rubber coated jaw faces for clamping of the material, which are fitted to a tensile testing machine - normally a single-column or dual-column table-top design. Using materials testing software, such as Bluehill® 2, you are able to input specimen details, set the desired test control, automatically calculate the desired results and statistics, and produce a test report all in accordance with the standard.

We suggest reviewing EN868-5 to fully understand the test fixture and results requirements. 


About This Solution

Related Standards: EN 868-5
Specimen Type: Irregular | Medical Adhesives
Materials: Packaging
Type of Test: Peel, Tear and Friction | Tension (What is a Tension Test?)
Business Sector: Biomedical/Medical/Healthcare

You are viewing a static image in place of an informative flash animation. Please download Adobe® Flash Player and refresh this page to view the animation.
Interested in this Solution?
Contact Us

Related Literature:

2710-100 Series   2710-100 Series
2712 Series Pneumatic Action Grips   2712 Series Pneumatic Action Grips
3300 Brochure    3300 Brochure
5900 Series General Brochure   5900 Series General Brochure
Bluehill­® 2 Brochure   Bluehill­® 2 Brochure
EN 868-5:1999   EN 868-5:1999

Our Labs

Instron application labs are used for both research and development, as well as testing customer samples to assist in finding the right equipment.
Contact our Application Specialists to determine the solution that's best for you!