Characterization and Fatigue of Intervertebral Disc Prostheses (ASTM F2346)

Intervertebral disc replacement is a surgical technique for the treatment of lower back pain related to degenerative disc disease. The advantage of this technique over traditional spinal fusion is that it preserves or restores motion in the spine and has the potential to delay the onset of degeneration of healthy discs at adjacent levels in the spine. Disc prostheses are designed to be load bearing over the physiological range of disc motion and to give years ofpain- and trouble-free operation in the body. Understanding the static and dynamic characteristics of a particulardevice allows manufacturers and designers to ensure their product is proven and accepted.

ASTM F2346 provides a methodology for characterising the static strength and dynamic fatigue behavior of disc prostheses. The rigorous testing regimes in this standard aim to scientifically validate any prosthesis design.

With a typical test run lasting for 10 million cycles and requiring both axial and torsional loading, it is vital that a testing system copes with these performance demands and delivers the highest quality of results. The requirement to conduct these tests in a wet environment adds to the complexity of the system.

An 8874 axial-torsional system, with the addition of a temperature controller and re-circulator unit, allows device manufacturers and contract research laboratories to conduct both static and cyclic testing on a range of implant designs. The bath, in which saline flow and temperature are controlled, provides a stable environment.

With the use of specialized test fixturing, the 8874 system’s combined axial-torsional actuator allows for characterization to be conducted in axial compression, compression-shear, and compression-torsion test modes for both articulating discs of traditional metal-on-metal, or metal-on-polyethylene design. It’s also used for the next generation of protheses, which feature an elastomeric component to give axial compliance under load that mimic the biomechanics of the natural disc. The challenges of developing testing equipment and fixtures that allow for unconstrained six degrees of freedom of motion and satisfy demanding ASTM requirements have been overcome by the 8874 installation at Ranier Technology Limited, Cambridge, UK. It is used for both performance and fatigue characterization of their elastomeric lumbar disc prosthesis – CAdisc-L.

We recommend that you review the standard to fully understand its requirements.

Are you interested in other types of test on spinal implants and spinal devices? Details can be found here.


About This Solution

Related Standards: ASTM F2346
Specimen Type: Implant
Materials: Biomedical | Biomaterials
Type of Test: High-Cycle Fatigue | Torsion | Shear
Business Sector: Biomedical/Medical/Healthcare
Credits: This application note was developed with assistance and photos from Ranier Technology Limited
8874 Axial-Torsion System with Test Fixtures to Meet ASTM 2346

CAdisc-L Lumbar Prosthesis, Courtesy of Rainer Technology Limited
You are viewing a static image in place of an informative flash animation. Please download Adobe® Flash Player and refresh this page to view the animation.
Interested in this Solution?
Contact Us

Related Literature:

Solutions for Biomedical Testing   Solutions for Biomedical Testing

Our Labs

Instron application labs are used for both research and development, as well as testing customer samples to assist in finding the right equipment.
Contact our Application Specialists to determine the solution that's best for you!