Flexure Test Typical Configurations
|
|
|
|
Rigid Plastic 3-Point Flex Fixture |
Rigid Plastic 4-Point Flex Fixture |
|
|
Select image to enlarge
|
Select image to enlarge
|
|
Test Standard |
ASTM D 790 |
ASTM D6272 |
|
Load Range |
100 lb |
200 lb |
|
Strain Range |
5% Flexural Strain |
|
Test Speed |
0.05 to 0.1 in/min |
|
Fixture |
3-Point Flex Fixture with 5mm Radius Anvils |
4-Point Flex Fixture with 5mm Radius Anvils and Deflectometer |
|
Fixture Span to Specimen Depth |
16:1 (can be as high as 60:1 for composite materials) |
|
Inner Span (Upper Anvils) |
N/A |
Set to ½ the Outer Span (Lower Anvils) |
|
Extensometer |
N/A |
Yes, Mounts on the Deflectometer |
|
Frame |
200 lb to 1000 lb |
|
Load Cell |
100 lb |
200 lb |
|
Common Specimen Size |
0.5" wide x 0.125" thick x 5.0" long |
|
Applications |
Polyethylene, Polypropylene, Acrylic, Polycarbonate, Polyester, ABS, Nylon, Acetates |
|
Industries |
Automotive, Aerospace, Consumer Products, Packaging, Biomedical, Electronics |
|
Related Information
|